Have a question? It may have already been answered. Find answers to support
questions often asked by picking a category from the drop down menu below.
Can't find an answer to your question here? Ask
us a new one. One of the
experts from our technical support staff will be happy to promptly respond.
You can also call 888-SEACHEM if you would like to speak directly with
a a technical support representative, or check our discussions
board.
MultiTest Nitrite & Nitrate
Q: I keep getting high nitrite readings, what could I be doing wrong?
A:
The test well you are using may be contaminated with Nitrate Reagent#2.
Soak the test well sheet in vinegar overnight, then rinse thoroughly
with tap water, followed by DI water, and dry completely.
Q: The nitrite in the tank is reading approximately 10 ppm. Can I still
get an accurate reading of nitrate?
A: If you are getting a pronounced
nitrite reading, it is not possible to get an accurate nitrate
reading, since the nitrite reading will interfere. You should be able to
make a very approximate assessment of nitrate, since the nitrate color will
be additive to the nitrite color.
Q: The nitrate reading on my kit is reading too low. I checked the reference
and it looks lower than it should be?
A: There was an unfortunate
typo on several batches of the Nitrate instructions in which the
reference was listed as 1 mg/L; it is actually 10 mg/L (which actually makes
your problem even more pronounced!). However, if you do the two following
two things your test should read the reference correctly (a) use 2 full
stems of sample instead of 1 and (b) vigorously shake, squeeze and tap on
a table the Nitrate Reagent#2 so that you obtain a very distinctly grey
drop. Assuming you get a correct reading with the reference, you can now
run your test samples in the same manner and be assured of the test's accuracy.
If have followed the above instructions yet you still get too low a reading
for the reference, then contact us to obtain replacement reagents.